ClearanceBrief
Recall Z-2188-2019 ยท JAK System, X-Ray, Tomography, Computed

Philips Medical Systems (Cleveland) Inc

Ingenuity Core Model # 728321, computed tomography x-ray system

DateAug 6, 2019 (initiated Jun 24, 2019)
StatusTerminated
ReasonIn certain instances when performing a cardiac step and shoot acquisition with phase tolerance selected, images may be labeled as if phase tolerance had been applied when phase tolerance had not been applied
Root causeOther
ActionPhilips Healthcare issued a Field Safety Notification (FSN)72800720/88200522 1. Identify affected product 2. There are 3 scenarios users may experience which result in identical images that are incorrectly labeled as phase tolerance images. A. When performing a Step and Shoot acquisition with phase tolerance selected, images may be reconstructed identically (a single phase), but are labeled as different phases. This occurs when Start Final Recon is selected prior to display of the ECG wave at the top of the acquisition window. B. When performing a Step and Shoot acquisition with phase tol
Quantity202
DistributionWorldwide distribution. US nationwide including Puerto Rico, Canada, Mexico, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia, Cote D'Ivoir
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K160743Philips Ingenuity CT
Philips Medical Systems (Cleveland), Inc., Aug 8, 2016