Recall Z-2294-2019 ยท IOR Wheelchair, Mechanical

TiSport, Llc

TiLite X-Short anti-tip tubes, models TRA, AERO T, ZRA, AERO Z, 2GX, AERO X, Twist, and AERO R, a component of the TiLite wheelchair (anti-tip extension for manual wheelchairs). Product Usage: The affected component is part of a manual wheelchair to provide mobility to physically impaired individuals. The affected component is used to prevent the wheelchair from tipping backwards.

DateSep 11, 2018
StatusTerminated
ReasonThe anti-tip tubes potentially fail in certain situations and the wheelchairs can tip backwards.
Root causeComponent design/selection
ActionThe recalling firm issued distributor letters dated 9/7/2018 via certified mail beginning on 9/11/2018. The letter identified the affected product, problem and actions to be taken.
Quantity1,452 wheelchairs
DistributionDistribution was nationwide, including Puerto Rico. There was also government distribution but no military distribution. Foreign distribution was made to Canada, Australia, Brazil, China, Italy, Saudi Arabia, Switzerland, United Kingdom, and Uruguay.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K023606TILITE YG AND TILITE YGS
Tisport, Nov 19, 2002
K072311TILITE AERO X
Tisport, LLC, Sep 21, 2007
K990358CROSS SPORT, RIGID WHEELCHAIR
Tisport, Feb 26, 1999