ClearanceBrief
Recall Z-2082-2019 ยท IYE Accelerator, Linear, Medical

Medical Intelligence Medizintechnik Gmbh

IGuide System, software versions: 2.2.0, 2.2.1, 2.2.2 Product Usage: Linear Accelerator. The intended use of the device is the control of accurate patient positioning with assistance of a 30 Tracking System in a radiotherapy environment.

DateJul 18, 2019
StatusTerminated
ReasonSometimes, although the HexaPOD has reached the target position, the Constant Correlation Check (CCC) can be too sensitive and iGUIDE may incorrectly demand a verification scan.
Root causeSoftware design
ActionThe firm sent an Urgent Important Field Safety Notification letter dated July 2019 via email. The identified the affected product, problem and actions to be taken. Customers were instructed to post this notice in a place accessible to all users, e.g. Instructions for Use, until this action is closed and advise the appropriate personnel, working with this product, on the content of this letter.
Quantity496 units
DistributionWorldwide Distribution - US Nationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K072079IGUIDE SYSTEM
Medical Intelligence Medizintechnik-GmbH-An Elekta, Aug 14, 2007