ClearanceBrief
Recall Z-2236-2019 ยท JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented

Zimmer Biomet, Inc.

Orthopedic Salvage System (OSS) Hybrid Poly Tibia Item # 151815

DateJul 2, 2019
StatusTerminated
ReasonCleaning processes potentially being ineffective
Root causeMaterial/Component Contamination
ActionRisk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. Complete Certificate of Acknowledgement and send to Corporate Quality. PostMarket@zimmerbiomet.com. This form must be returned even if you do not have affected products at your facility. 4. Retain a copy of the acknowledgement form with your recall
Quantity19
DistributionDistributed to the following states in the USA: AZ, CA, DC, GA, IL, KY, LA, MA, MD, MN, NC, NJ, NM, OH,OK, OR, SC, TX. Distributed to Japan and Netherlands.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K142814Biomet Orthopaedic Salvage System (OSSTM) - Proximal Femoral & Hybrid Tibial
Biomet Manufacturing, LLC, Jun 25, 2015