Recall Z-2320-2019 ยท CEM Electrode, Ion Specific, Potassium

Radiometer America Inc

ABL90 FLEX Analyzer REF 393090 UDI:05700693930909

DateAug 22, 2019 (initiated Jul 10, 2019)
StatusTerminated
ReasonSoftware Security; The action is being initiated because of software security vulnerabilities with the firm's analyzer operating system, which may cause the device to shutdown or reboot resulting in delayed medical treatment.
Root causeSoftware design
ActionOn (date), the firm will email a "Urgent Field Safety Notice" to all customers and will follow-up by mailing a copy of the "Urgent Field Safety Notice" to all customers via Federal Express with tracking and proof of delivery. The Urgent Field Safety Notice ask customers to do the following: 1. Check which version of the operating system (OS) is currently installed on your analyzer: o Logon to the analyzer o Touch Menu, Utilities, Temporary shutdown, and then Confirm o During the shutdown process the analyzer displays the version of OS installed " If the OS version is one of the following
Quantity13,042 units
DistributionUS: AB, AL, AR, BC, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NY, OH, OK, OR, PA, SC, TX, VA, VT, WA, WI, WV OUS: Canada, Netherland, China, Czech Republic, Denmark, Germany, Spain, France, Hugary, , indiat Japan, South Korea, United Kingdom, Australia
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K092686ABL90 FLEX SERIES
Radiometer Medical Aps, Aug 6, 2010