Recall Z-2156-2019 ยท LKD Monitor, Carbon-Dioxide, Cutaneous

SenTec AG

SenTec Membrane Changer Insert [5 pcs - Product Usage: is indicated for continuous, non-invasive monitoring of carbon dioxide tension and oxygen tension as well as oxygen saturation and pulse rate in adult and pediatric patients. In neonatal patients the SDMS is indicated for carbon dioxide and oxygen tension monitoring only. Oxygen tension monitoring is contraindicated for patients under gas anes

DateMay 2, 2019
StatusTerminated
ReasonThere is an increased chance (approx. 5%) of no electrolyte or a reduced amount of electrolyte being applied to the sensor during the membrane change procedure.
Root causeProcess control
ActionOn May 2, 2019, the firm sent Guidance for SenTec Distributors to their distributors advising them of the issue, requesting them to check their stock and notify their end users via a voluntarily initiated field correction action associated with Membrane Changers and their Inserts which was enclosed with the "recall" letters. The distributors were to report back to SenTec via email.
Quantity205 pcs
DistributionWorldwide distribution - US Nationwide in the state of Missouri and countries of Argentina, Australia, Austria, Belgium, Chile, China, Columbia, Costa Rica, Czech Republic, Denmark, Ecuador, Europe, France, Germany, Greece, Guatemala, Hong Kong, Hungary, Indonesia, Israel, Italy, Kuwait, Malaysia, F
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041548SENTEC DIGITAL MONITOR MODEL #SDM
Sentec, Inc., Aug 9, 2004
K101690SENTEC DIGITAL MONITOR, V-SIGN SENSOR AND V-SIGN SENSOR 2 WITH DIGITAL SENSOR ADAPER CABLE
Sentec AG, Dec 3, 2010
K151329SenTec Digital Monitor, OxiVenT Sensor, V-Sign Sensor, Staysite Adhesive Pad, V-STATS PC Software including V-CareNet
Sentec AG, Dec 17, 2015