Recall Z-2228-2019 ยท OLO Orthopedic Stereotaxic Instrument
Intellijoint Surgical, Inc.
Workstation, G2, Configured (P/N 000-0241)
| Date | Aug 8, 2019 (initiated Jun 28, 2019) |
| Status | Terminated |
| Reason | Apple has determined that, in a limited number of older generation 15-inch MacBook Pro Units, the battery may overheat and pose a fire safety risk. To date, Intellijoint has distributed, 15-inch MacBook Pros which are the Workstation component (affected P/N 000-0241) in the Intellijoint HIP System. |
| Root cause | Nonconforming Material/Component |
| Action | The firm initiated the recall by email on 06/28/2019. The letter requested the distributor locate and return the MacBooks. |
| Quantity | 23 units |
| Distribution | IL, CA, NY, TX, RI, Australia |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K172849 | Intellijoint HIP System (Cart), Intellijoint HIP System (Portable), Intellijoint HIP Instrument Tray Intellijoint Surgical, Inc., Nov 16, 2017 |