Recall Z-0189-2020 ยท FPA Set, Administration, Intravascular

Smiths Medical ASD Inc.

GRIPPER Needles sold individually. Labeled as the following item descriptions: 1. NDL, GRIPPER, 22G X 1" (25MM), 12/BX; 2. NDL, GRIPPER, 22G X 1.25" (32MM), 12/BX; 3. NDL, GRIPPER, 20G X 1.25" (32MM), 12/BX; 4. NDL, GRIPPER, 19G X 1" (25MM), 12/BX; 5. NDL, GRIPPER, 19G X 1.25" (32MM), 12/BX; 6. NDL, GRIPPER, 22G X .75" (19MM), 12/BX; 7. NDL, GRIPPER, 19G X .75" (19MM), 12/BX;

DateJul 8, 2019
StatusTerminated
ReasonPotential for certain GRIPPER Needles to contain an occluded or blocked needle.
Root causeProcess control
ActionSmiths Medical notified customers during the week of 07/08/2019 via "URGENT MEDICAL DEVICE RECALL NOTICE" letter. The Recall Notice advised customers that Smiths Medical has initiated a voluntary field removal for specific lots of GRIPPER needles and specific lots of PORT-A-CATH Trays containing GRIPPER needles. Customers were instructed to examine inventory for affected product and complete and return the Response Form to GRIPPERocclusion2019@smiths-medical.com. Once the response form is received, Customer Service will provide a shipping label to return all affected product to Stericycle
Quantity3,952,425 devices total
DistributionNationwide distribution to AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, PR. International distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K870866THE GRIPPER(TM) PORT-A-CATH NEEDLES
Pharmacia, Inc., Jun 25, 1987