Recall Z-0163-2020 ยท HBL Holder, Head, Neurosurgical (Skull Clamp)

Deerfield Imaging, Inc.

IMRIS MR/X-ray head fixation device, Model HFD200

DateJun 26, 2019
StatusTerminated
ReasonBreaking of the HFD200 Drive Knob force indicator due to over-torqueing of the drive knob beyond the maximum allowable pinning force specification may result in the clinical staff unknowingly pinning the patient's skull with insufficient force; the clinical staff unknowingly pinning the patient's skull with excessive force; or the clinical staff identifying the issue and not being able to continue with the procedure.
Root causeDevice Design
ActionThe firm disseminated a field safety notice on 06/26/2019 by letter. The notice explained the problem and requested that the consignee inspect the skull clamp prior to use and be aware that overtightening of the drive knob may result in failure. The drive knobs will be replaced with an updated design when available.
Quantity3 units
DistributionTN, NH, FL
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K113748IMRIS MR/X-RAY HEAD FIXATION DEVICE (HFD)
Imris, Inc., Mar 23, 2012