Recall Z-1965-2019 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

GE Healthcare, LLC

Repair RIC5-9A-RS System, imaging, pulsed doppler, ultrasonic and Refurbish RIC-5-9-D REALTIME 4D

DateJul 12, 2019 (initiated Jul 2, 2019)
StatusTerminated
ReasonPotential for a transducer mis-alignment in certain transvaginal probes.
Root causeDevice Design
ActionGE Healthcare notified customers on about 07/02/2019 via "URGENT MEDICAL DEVICE CORRECTION" letter. Customers were informed that they could continue to use the affected probes, but prior to performing a biopsy, customers should make sure they follow the instructions for safe use in the Basic User Manual to ensure proper device alignment, specifically including the biopsy, biopsy safety and biopsy setup sections. GE Healthcare will correct all affected devices at no cost and a GE Healthcare representative will contact customers to arrange for the correction. Customers were also instruc
Quantity2 devices
DistributionDistribution in the US to Texas. International distribution to France, Greece and Switzerland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K180374Voluson S8/ Voluson S10/ Voluson S10 Expert
GE Healthcare, Mar 13, 2018
K180535Voluson P6 and Voluson P8
GE Healthcare, Mar 29, 2018
K181783LOGIQ P9; LOGIQ P7
GE Healthcare, Sep 20, 2018