Recall Z-2129-2019 ยท EAJ Apron, Leaded
Burlington Medical, LLC
Porta Shield, Product Number Porta Shield, Part Numbers PTS2, PTS3, PTS4 Product Usage: Products are designed to protect healthcare workers against scatter (secondary) radiation while performing fluoroscopic and radiographic procedures by reducing the amount of radiation reaching the user.
| Date | Jul 30, 2019 (initiated Jun 7, 2019) |
| Status | Terminated |
| Reason | The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions. |
| Root cause | Under Investigation by firm |
| Action | Urgent Medical Device Recall notification letters were sent to customers on 6/7/19. Please take the following actions: Cease using recalled products immediately, quarantine the products, and prepare the products for return shipment to Burlington Medical. Contact Burlington Medical at (800) 221-3466 for specific return instructions. Note that Burlington Medical Customer Care will work with you to facilitate the return of recalled products, including by issuing you a Recall Tracking number. Burlington Medical will replace all recalled products at no cost to you. The replacement product will |
| Quantity | 9 |
| Distribution | US Nationwide distribution. The products were distributed to the following foreign countries: Korea, Thailand. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.