Recall Z-0353-2020 ยท LGX Table, Examination, Medical, Powered

Heritage Medical Products, Inc.

2000 Vascular with 30 Degree option

DateSep 2, 2016
StatusTerminated
ReasonPotential for bed to drop suddenly at the head while a patient is laying on it.
Root causeDevice Design
ActionHeritage Medical Products, Inc. notified customers on about 09/02/2016 via "Advisory Notice." Customers were informed that Heritage Medical has designed a reinforcement kit as a correction, which will be supplied to customers. Customers were instructed to contact Heritage Medical, provide the serial number of their bed to ensure it is affected, and Heritage Medical will provide a free kit that will reinforce the area on the affected devices that has experienced failures. The contact information is a follows: Heritage Medical Products, Inc., ATTN.: Trunnion Reinforcement Kit, 10380 CR 6310,
Quantity167 beds (159 US, 8 OUS)
DistributionNationwide distribution to AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, OH, OK, OR, PA, SC, TX, UT, VA, WI, WY, PR. International distribution to Canada, Saudi Arabia, Trinidad & Tobago.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.