Recall Z-2230-2019 ยท IYE Accelerator, Linear, Medical

Elekta, Inc.

iGUIDE System with software build iGuide 2.2.0, 2.2.1, and 2.2.2

DateJun 19, 2019
StatusTerminated
ReasonPotential for iGUIDE to incorrectly monitor the 3D position.
Root causeSoftware Design Change
ActionElekta, Inc. notified customers on about 06/19/2019 via URGENT IMPORTANT FIELD SAFETY NOTIFICATION" letter. Customers were instructed to follow the below steps: 1. Log on to iGUIDE and move the 3D position. If the 3D position is equal to the PARK position, the system is not affected. 2. If the 3D position is not equal to the PARK position, log on to iGUIDE with Operator and Administrator rights. (If no user is available, go to the iGUIDE User Manager and add an appropriate user.). 3. Move the HexaPOD to the 3D Position (If 3D position is equal to the DRIVE position, both checkmarks wil
Quantity374 devices
DistributionNationwide distribution to AK, AL, CA, CO, CT, FL, GA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, VA, VT, WA, WI. Worldwide distribution to Algeria, Argentina, Australia, Austria, Belgium, Botswana, Canada, Chile, China, Colombia, Cyprus, Czec
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K072079IGUIDE SYSTEM
Medical Intelligence Medizintechnik-GmbH-An Elekta, Aug 14, 2007