Recall Z-1956-2019 ยท BTR Tube, Tracheal (W/Wo Connector)
Centurion Medical Products Corporation
Sheridan Endotracheal Tube contained inside Centurion kit code TC7855.
| Date | Jul 30, 2019 (initiated Jun 7, 2019) |
| Status | Terminated |
| Reason | Centurion is conducting a sub-recall for Teleflex Sheridan branded Endotracheal Tracheal Tubes due to increased incidence of Sheridan connector becoming disconnected from Endotracheal Tube. |
| Root cause | Device Design |
| Action | On June 7, 2019, Centurion notified Medline (the distributor) that they were conducting a recall via an Urgent Product Recall Notice. Medline was instructed to cease further distribution and to notify any customers who received this product. On June 14, 2019, Medline informed its customer of the recall via a letter titled "CENTURION MEDICAL PRODUCT SUB-RECALL*** IMMEDIATE ACTION REQUIRED". REQUIRED ACTION FOR CUSTOMERS: 1. Immediately check your stock for the affected item number and the affected lot number listed on the attachment. Quarantine all affected product. 2. Please return the |
| Quantity | 80 kits |
| Distribution | Distributed to one account in Arkansas. |
| FDA record | Open on fda.gov |