Recall Z-2009-2019 ยท KPS System, Tomography, Computed, Emission

Philips Medical Systems (Cleveland) Inc

BrightView SPECT, Model Number 882480

DateJun 10, 2019
StatusTerminated
ReasonAn issue with the handcontroller may result in either spontaneous uncommanded motions or continued motion after the buttons were released.
Root causeUnder Investigation by firm
ActionUrgent Field Safety Notice Medical Device Correction letters dated 6/4/19 were sent to customers. Discontinue Use of the BrightView Handcontroller. Remove the handcontroller from the charging station and store in a secure location inaccessible by technicians, at least 20 meters (65 feet) away from the system and outside the imaging room. If 20 meters (65 feet) is not possible, then store the handcontroller in an area that is shielded, such as a hot lab. Do not dispose of the hand controller, as it will be needed once a solution is identified. Instead of the physical handcontroller, use th
Quantity696
DistributionWorldwide Distribution - US Nationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K080927BRIGHTVIEW VCT IMAGING SYSTEM
Philips Medical Systems (Cleveland), Inc., Apr 11, 2008