Recall Z-2151-2019 ยท FKX System, Peritoneal, Automatic Delivery
Fresenius Medical Care Renal Therapies Group, LLC
Liberty Cycler Cassettes (a disposable cassette that is intended to be used to perform peritoneal dialysis (APD) in conjunction with the Liberty cycler as follows: (1) LIBERTY Line Cycler Set with Stay-safe connectors (Dual Patient Connector) Part Number: 050-87212 (2) LIBERTY Line Cycler Set with Stay-safe connectors (Single Patient Connector) Part Number: 050-87215 (3) LIBERTY Integrated Cycler
| Date | May 29, 2019 |
| Status | Terminated |
| Reason | Additional Instructions: step-by step handling and inserting the cassette into the Liberty Cycler set. A damaged cassette can cause peritoneal dialysis (PD) solution to leak into the cycler and is a potential risk for contamination. |
| Root cause | Use error |
| Action | Fresenius issued on 05/29/19 The IMPORTANT MEDICAL DEVICE INFORMATION notification explains to the Home Therapy Nurses/Clinics to review the new flyer with each of their APD patients who use a Liberty Cycler. For the patient notification, it will encourage them to use and follow the attached instructional flyer during set up for their daily dialysis treatment. The notification will have a copy of the new step-by-step instructional flyer. Included is a Reply Form for confirmation the consignee read and understands the notice sent to them. |
| Quantity | 33732973 Cassettes; 18,674 Liberty Cyclers |
| Distribution | US Nationwide Distribution |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K173718 | Liberty Cycler Set Fresenius Medical Care Renal Therapies Group, LLC, May 1, 2018 |