Recall Z-2763-2019 ยท CGR Radioimmunoassay, Cortisol

Beckman Coulter Inc.

The Access Cortisol Reagent. REF (Catalog Number) 33600, Lot 831755. A paramagnetic particle, chemiluminscent immunoassay. Product Usage: The assay is a paramagnetic particle, chemiluminescent immunoassay for the quanitative determination of cortisol levels in human serum, plasma (heparin, EDTA) and urine using the Access Immunoassay Systems.

DateSep 18, 2018
StatusTerminated
ReasonCross contamination may have occurred between wells for the reagent pack. As a result of this defect, patient results may be negatively impacted (falsely decreased) by up to 16%).
Root causeProcess design
ActionAn Urgent Medical Device Recall Letter was sent during the week of 9/17/2018 to all affected customers via mail and or email. Customers were instructed to discontinue use and discard remaining product. Customers were also instructed to consider a review of the Cortisol results generated with the affected lot number at the discretion of the Medical Director.
QuantityAccess Cortisol Reagent Assay
DistributionWorldwide distribution - US Nationwide and the countries of Canada, Lithuania, Mexico, Panama, Serbia, South Africa, and Vietnam.