Recall Z-2057-2019 ยท HGI Electrocautery, Gynecologic (And Accessories)

CooperSurgical, Inc.

LEEP Precision Integrated System 120V & LEEP Precision Generator, Model Nos. LP-10-120 & LP-20-120

DateJun 18, 2019
StatusTerminated
ReasonThe products' cut, coagulate, or blend function may not operate with the use of the Foot Pedal.
Root causeDevice Design
ActionThe firm notified customers of the recall via Urgent Medical Device Recall letters dated June 18, 2019. Customers were informed that the device may fail to cut, coagulate, or blend tissue with use of the Foot Pedal. Serious injuries could occur due to this failure mode. Customers were instructed to do the following: - Use the Hand-Switch Pencil and Cord (Part Numbers: 395310/LP50101) to activate the cut, coagulation, or blend functions by hand control instead of foot action using the Foot Pedal. - Complete the acknowledgement and receipt form and return to the firm. If you have additional
Quantity358
DistributionNationwide domestic distribution. International distribution to Canada and Costa Rica.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K963348WALLACH INTEGRATION UNIT
Wallach Surgical Devices, Inc., Mar 19, 1997
K963653WALLACH QUANTUM 2000 ELECTROSURGICAL SYSTEM
Wallach Surgical Devices, Inc., Jun 11, 1997