Recall Z-2633-2019 ยท DQY Catheter, Percutaneous
Medtronic Vascular
SHERPA NX ACTIVE GUIDING CATHETER, 6F EBU4.5, .070", REF SA6EBU45. for cardiovascular use
| Date | Oct 1, 2019 (initiated Mar 15, 2019) |
| Status | Terminated |
| Reason | There is a potential for extensive loss of primary segment material exposing underlying stainless-steel braid wires on a subset of 6F Sherpa catheters. |
| Root cause | Component design/selection |
| Action | Medtronic initiated immediate communications to consignees on 15-Mar-2019. Consignees were asked to quarantine and/or return unused affected product, and confirmation of the immediate notification was requested per local requirements. In the US, Medtronic representatives notified consignees. Consignees were asked to return a Customer Confirmation Certificate. On 11-Apr-2019, Medtronic initiated the US follow up communications via hand-delivery and UPS 2-day delivery mailings to the March 2019 notification consignees and risk managers identified as part of the April 2019 scope expansion. C |
| Quantity | 496 units |
| Distribution | Worldwide distribution to US, Austria, Belgium, Bulgaria, Czech Republic, France, Germany, Greece, Hong Kong, Ireland, Italy, Kazakhstan, Luxembourg, Moldova, Netherlands, Poland, Portugal, Russian Federation, Saudi Arabia, Slovakia, South Africa, Spain, Sweden, Taiwan, and United Kingdom |
| FDA record | Open on fda.gov |