ClearanceBrief
Recall Z-1904-2019 ยท OWB Interventional Fluoroscopic X-Ray System

Siemens Medical Solutions USA, Inc

Siemens Artis Q.zen biplane-Interventional Fluoroscopic X-Ray System Model Number: 10848355

DateJun 6, 2019
StatusTerminated
ReasonArtis Zee, Artis Q, Artis Pheno or Artis Icono systems equipped with a DCS Extended or DCS Movable, the potential exists for individual screws of the screwed connection joining the pivot bearing to the ceiling-mounted support to work loose or snap off, can result in severe injuries to patients and operating personnel
Root causeDevice Design
ActionSiemens issued letter dated 6/6/19 advising of the problem, health risk and action to take: .Siemens is implementing a corrective action to all potentially affected users via Update Instruction AX023/18/S. Siemens Service organization will replace the screwed connection of the ceiling support according to specifications / instructions. In the event the bolt connection cannot be replaced due to technical conditions, the complete bearing will be replaced.
Quantity11
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181407Artis zee/zeego & Artis Q/Q.zen
Siemens Medical Solution USA, Inc., Aug 15, 2018