Recall Z-1901-2019 ยท OWB Interventional Fluoroscopic X-Ray System
Siemens Medical Solutions USA, Inc
Siemens Artis Q zeego ---Interventional Fluoroscopic X-Ray System Model Number: 10848283
| Date | Jun 6, 2019 |
| Status | Terminated |
| Reason | Artis Zee, Artis Q, Artis Pheno or Artis Icono systems equipped with a DCS Extended or DCS Movable, the potential exists for individual screws of the screwed connection joining the pivot bearing to the ceiling-mounted support to work loose or snap off, can result in severe injuries to patients and operating personnel |
| Root cause | Device Design |
| Action | Siemens issued letter dated 6/6/19 advising of the problem, health risk and action to take: .Siemens is implementing a corrective action to all potentially affected users via Update Instruction AX023/18/S. Siemens Service organization will replace the screwed connection of the ceiling support according to specifications / instructions. In the event the bolt connection cannot be replaced due to technical conditions, the complete bearing will be replaced. |
| Quantity | 2 |
| Distribution | Nationwide |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K181407 | Artis zee/zeego & Artis Q/Q.zen Siemens Medical Solution USA, Inc., Aug 15, 2018 |