Recall Z-2078-2019 ยท JIT Calibrator, Secondary

Abbott Laboratories

ARCHITECT BNP Calibrators Ptoduct Usage: The ARCHITECT BNP Calibrators are for the calibration of the ARCHITECT iSystem when used for the quantitative determination of human Btype natriuretic peptide (BNP) in human EDTA plasma.

DateMay 24, 2019
StatusTerminated
ReasonPresence of a time dependent, stability drift in patient and control results returned from ARCHITECT BNP testing.
Root causeUnder Investigation by firm
ActionAbbott Laboratories notified customers on about 05/24/2019 via "Product Recall" letter. If customers did not have an alternate calibrator and/or control lots available in inventory and a valid calibration calibration curve has been generated, then customers should immediately order replacement calibrator and/or control lots. Valid calibration curves generated with the affected calibrator lots can still be used as long as controls not listed on the customer notification letter remain within range. The control lots listed in the letter CANNOT be used to validate the calibration curves. Destroy
Quantity5,483 units
DistributionWorldwide Distribution - US Nationwide in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. Distributed worldwide to ALGERIA, ANDORRA, AUSTRIA,