ClearanceBrief
Recall Z-1750-2019 ยท KPS System, Tomography, Computed, Emission

GE Healthcare, LLC

Discovery Ml DR Product Usage: The systems are intended for head and whole body attenuation corrected Positron Emission Tomography (PET) imaging and localization of emission activity in patient anatomy by means of integrated PET and CT images.

DateJun 19, 2019 (initiated Feb 12, 2019)
StatusOpen, Classified
ReasonGE Healthcare PET/CT Table Stop Without Stopping X-Ray Exposure
Root causeRadiation Control for Health and Safety Act
ActionGE Healthcare sent a notification letter to affected customer (Hospital Administrators/Risk Managers. Managers of Radiology/Cardiology, Radiologist/Cardiologist). The letter identified the affected product, problem and actions to be taken. Customers were informed GE Healthcare will without charge, remedy the defect or bring the product into compliance. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Quantity1113 in total
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133657DISCOVERY PET/CT 710
Ge Medical Systems, LLC, Mar 21, 2014