Recall Z-2359-2019 ยท DOP Mass Spectrometer, Clinical Use

Agilent Technologies, Inc.

MassHunter Quantitative Analysis Software, Version B.07.01, used with Triple Quadrupole Mass Spectrometers, models K6460 and K6420 Product Usage: A mass spectrometer for clinical use is a device intended to identify inorganic or organic compounds (e.g., lead, mercury, and drugs) in human specimens by ionizing the compound under investigation and separating the resulting ions by means of an elect

DateAug 15, 2018
StatusTerminated
ReasonWhen utilizing specific software versions of an LC/MS device, and Batch at a Glance in Compound Table view, a defect occurs when a new sample(s) is inserted. This defect creates a mismatch between the sample name and the column header whereby the sample header will be offset by one sample. The recalling firm requests that you discontinue using this workflow to generate reports because the Quant batch table will not display fully analyzed results or save any changes made.
Root causeSoftware design
ActionOn 08/13/18, Urgent Medical Device Recall notices were sent to customers informing them that the defect has been corrected with software version B.08.01. The recalling firm will contact customers to schedule the new software installation. Software version B.08.01 is being released in conjunction with software version B.08.03. This software suite includes improvements to other non-patient safety related defects such as the following fixes in Acquisition B.08.03. -Improvements to DMRM method setup which will prevent exceeding transition limits ("Failed to WriteScan" er
Quantity235
DistributionUS Nationwide. distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.