Recall Z-1886-2019 ยท LLZ System, Image Processing, Radiological

GE Healthcare, LLC

Centricity Universal Viewer with PACS-IW foundation 6.0, Model number 2090255-001 Product Usage: Centricity PACS-IW with Universal Viewer is a device that displays medical images (including mammograms) and data from various imaging sources. Images and data can be viewed, communicated, processed and displayed within the system or across computer networks at distributed locations. Lossy compressed

DateMay 17, 2019
StatusTerminated
ReasonThere is a potential that one or more images or image series may be missing from exams without a warning displayed in the viewer.
Root causeSoftware design
ActionThe firm issued a medical device correction notice on 05/17/2019. The letter stated the safety issue and provided additional instructions on the use of the system until a service engineer installs the correction. A firm representative will contact the consignee to arrange for the installation.
Quantity292 units
DistributionWorldwide Distribution - US Nationwide Austria, Belgium, Brazil Canada Colombia France Germany Ghana India Ireland Israel Italy Kuwait Mexico Netherlands Panama Poland Qatar Russia Saudi Arabia Slovenia South Africa Spain Switzerland Turkey United Arab Emirates United Kingdom Uruguay
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K150420Centricity Universal Viewer
GE Healthcare, May 26, 2015