Recall Z-1887-2019 ยท OCL Surgical Device, For Cutting, Coagulation, And/Or Ablation Of Tissue, Including Cardiac Tissue

AtriCure, Inc.

EPi-Sense Guided Coagulation System with Visitrax, MOdel No. CDK-1413 - Product Usage: The EPi-Sense Guided Coagulation System with VisiTrax, consists of a sterile, single-use, disposable coagulation electrode device (1cm, 2cm, and 3cm sizes provided) intended to be used to coagulate cardiac tissue. The flexible, cooled electrode device, with a suction stabilizer feature, transmits radiofrequency

DateMay 29, 2019
StatusTerminated
ReasonThe sterile package seal for the Epi-Sense Guided Coagulation System may be compromised, and has a worst-case reasonable harm of infection if used.
Root causeProcess control
ActionOn May 29, 2019, the firm notified customers of the recall via Urgent Advisory Notice. Customers were asked to take the following actions: " Immediately quarantine any affected product as identified within this Advisory Notice. " Contact AtriCure product complaints by phone at 1-866-349-2342 (select option 6) or e-mail to pcomplaints@atricure.com to request a Return Goods Authorization (RGA). " Return the attached Acknowledgement Form. See Attachment B. The acknowledgement form must be completed and returned even if product is not on hand by selecting the second option. " Adverse reactions
Quantity497
DistributionWorldwide distribution - US Nationwide distribution, and countries of Germany and Netherlands.