Recall Z-1781-2019 ยท LLZ System, Image Processing, Radiological

GE Healthcare, LLC

GE Centricity PACS-IW versions 3.6.0 through 3.7.3.9 SP1, SP2, SP3 and 3.7.3 SPa10, Model Numbers: a) 2052831-00x b) 2050578-0xx c) 2090228-002 d) 2052782-00x

DateMay 22, 2019
StatusTerminated
ReasonThere is a potential that one or more image series (i.e., all images within an image set) may be missing from an exam without a warning displayed in the viewer.
Root causeSoftware design
ActionThe firm initiated their medical device correction by letter on 05/2/2019. The letter explained the problem and the safety issue and provided the following safety instructions: 1) Process Root Causes: Investigation determined that the original PACS-IW foundation design did not incorporate adequate risk mitigation for this part of the image ingestion process. DFMEA of the image ingestion wing to wing process was completed as part of the investigation, and no additional areas where further mitigations are required were identified. a) If a discrepancy is identified, attempt to retransmit the
Quantity620 units
DistributionWorldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082318CENTRICITY PACS IW
Ge Healthcare Dynamic Imaging Solutions, Aug 25, 2008