Recall Z-1932-2019 ยท OXO Image-Intensified Fluoroscopic X-Ray System, Mobile
Medtronic Navigation, Inc.-Littleton
Medtronic 0-arm TM 1000 Imaging Systems: Bl-700-00027-XXX and B1-700-00028-XXX (where XXX indicates the system voltage): Product Name Model#/ CFN OARM ASSY 8170000027 SYSTEM PRODUCT BI70000027 OARM ASSY BI70000027R SYS PRODUCT RWK BI70000027R BASE OARM 8170000027100 SYSTEM !DOV BI70000027100 BASE OARM Bl70000027100R SYSTEM lOOV RWK Bl70000027100R BASE OARM 8170000027120 SYSTEM 120V 8170000027120 B
| Date | May 1, 2019 |
| Status | Terminated |
| Reason | The 0-arm 1000 Imaging System uses energy from the batteries to generate X-rays and move the system. If a charger board is not functioning properly, the batteries will not receive a full recharge of the potential capacity and results in the inability of the system to take X-rays, open the gantry, and move the system from one place to another, which may affect the ability to continue use within the operating room until the charger board and/or the batteries are replaced |
| Root cause | Device Design |
| Action | Medtronic Navigation issued letter May l, 2019 via FedEx Express Overnight delivery identifying the issue, health risk and action to take: provided a visual mitigation card to attach to the Image Acquisition System (IAS) of their affected systems to serve as a visual reminder. " Consignees will be asked to sign and return the consignee response form; acknowledging receipt of this information and confirmation of having attached the visual mitigation card to their affected system. Review and retain Attachment 8 for further guidance . Attachment B provides instructions for inspecting the batte |
| Quantity | 832 units |
| Distribution | Worldwide Distribution - US Nationwide Foreign: ALGERIA AUSTRALIA AUSTRIA BAHRAIN BELGIUM BRAZIL BULGARIA CANADA CANADA CHILE CHINA COLOMBIA CROATIA CZECH REPUBLIC DENMARK EGYPT FINLAND FRANCE GERMANY GREECE HONG KONG INDIA IRELAND ISRAEL ITALY JAPAN JORDAN KAZAKHSTAN KOREA KOREA, REPUBLIC OF KUWAIT |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K092564 | O-ARM IMAGING SYSTEM Medtronic Navigation, Inc. (Littleton), Jun 17, 2010 |