Recall Z-1914-2019 ยท CGS Nad Reduction/Nadh Oxidation, Cpk Or Isoenzymes

Siemens Healthcare Diagnostics, Inc.

Atellica CH Creatine Kinase (CK_L) Reagent, Product Code 11097640, UDI Number: 00630414006642 - Product Usage: The ADVIA Chemistry Creatine Kinase (CK_L) assay is for in vitro diagnostic use in the quantitative determination of creatine kinase activity in human plasma (lithium heparin) or serum on ADVIA Chemistry XPT systems. The assay can be used to aid in the diagnosis and treatment of myocardia

DateMay 3, 2019
StatusTerminated
ReasonA steady upward trend in blank (u) absorbance was observed
Root causeNonconforming Material/Component
ActionSiemens Healthineers issued Urgent Medical Device Recall notices to domestic customers and Urgent Field Safety Notices to international customers dated May 8, 2019. Actions to be Taken by the Customer 1) Discontinue use of and discard the kit lots listed on the notice. 2) Review your inventory of these products to determine your laboratorys replacement needs and provide information to Siemens for reporting to the authorities. 3) Complete and return the Field Correction Effectiveness Check letter within 30 days. 4) Please review the letter with your Medical Director. 5) If you have received
Quantity899
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K160724ADVIA Chemistry Creatine Kinase (CK_L) Assay, ADVIA Chemistry Enzyme 3 Calibrator
Siemens Healthcare Diagnostics, Inc., Sep 7, 2016