Recall Z-1857-2019 ยท FTD Lamp, Surgical
Maquet Cardiovascular Us Sales, Llc
Maquet VOLISTA StandOP Surgical Lights- Product Code: ARDVST229000A Model: VST60SF AIM
| Date | Jun 21, 2019 (initiated May 28, 2019) |
| Status | Terminated |
| Reason | Potential that the coating of the light head forks on the VOLISTA Surgical Lights may crack, posing the risk of particles falling into the surgical field during surgery that may result in patient infection |
| Root cause | Device Design |
| Action | Getinge/Maquet SAS issued on 5/28/19 Urgent Medical Correction letter via FedEx 2 day Delivery with Signature Proof of Delivery (SPOD). Letter provides identification of issue, health risk and action to take: Before using the VOLISTA StandOP, please inspect the forks and check if cracks on the coating are visible as instructed in the VOLISTA StandOP manual; If a coating crack is detected, please STOP USING the VOLISTA StandOP and contact your Getinge representative. Complete and sign the attached MEDICAL DEVICE CORRECTION RESPONSE FORM . Return the completed form to Maquet/Getinge by e-maili |
| Quantity | 2 units |
| Distribution | Worldwide distribution. US Nationwide, ALBANIA GHANA PORTUGAL ARGENTINA HONG KONG ROMANIA AUSTRIA HUNGARY RUSSIA AUSTRALIA INDIA SAUDI ARABIA AZERBAIJAN IRAN SINGAPORE BAHAMAS IRELAND SLOVAKIA BANGLADESH IRAQ SLOVENIA BELGIUM ISRAEL SOUTH AFRICA BOLIVIA ITALY SOUTH KOREA BRAZIL JAPAN SPAIN BULGARIA |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K130513 | VOLISTA 400; VOLISTA 600 Maquet S.A.S., May 8, 2013 |