ClearanceBrief
Recall Z-1829-2019 ยท GDW Staple, Implantable

Covidien LLC

EGIA 60 ARTICULATING MED THICK SULU , EGIA60AMT

DateMay 17, 2019
StatusOpen, Classified
ReasonThe device may be missing one of two pin components that maintain alignment of the device jaws, resulting in incomplete staple formation which may lead to bleeding, anastomotic leak, peritonitis, or pneumothorax which can result in the potential for infection and/or sepsis.
Root causeProcess control
ActionAll consignees were notified by letter delivered via Federal Express and/or certified mail beginning May 17, 2019. The letter informs customers of the recall of specific item codes and production lots of Covidien Endo GIA Articulating Reloads with Tri-Staple Technology due to the potential for a device to be missing one of two pin components that maintain alignment of the device jaws. Through this notification customers are instructed to take the following actions: " Immediately quarantine and discontinue use of the affected item codes and lots. " Return affected product. " Forward the informa
Quantity3,113,280
DistributionNationwide domestic distribution, worldwide foreign distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K111825ENDO GIA(TM) STAPLERS, DST SERIES(TM) GIA(TM) STAPLERS, DST SERIES(TM) TA(TM) STAPLERS
Covidien, Formerly US Surgical A Divison of Tyco H, Oct 25, 2011
K160176Signia Stapler
Covidien, Apr 26, 2016