Recall Z-2296-2019 ยท JKA Tubes, Vials, Systems, Serum Separators, Blood Collection
Becton Dickinson & Company
BD Vacutainer Lithium HeparinN (LH) 37 USP Units Blood Collection Tubes, Catalog Number 366664
| Date | Jun 12, 2019 |
| Status | Terminated |
| Reason | False elevation of carboxyhemoglobin (COHb) results have been reported when using venous blood gas samples collected with tubes analyzed with the IL GEM 4000 instrument. |
| Root cause | Device Design |
| Action | Urgent Medical Device Correction notification letters dated 6/12/19 were sent to customers. |
| Quantity | 1,436,228,004 total |
| Distribution | Worldwide Distribution, including US Nationwide |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K901449 | VACUTAINER PLUS WITH HEMOGARD CLOSURE Bd Becton Dickinson Vacutainer Systems Preanalytic, Aug 9, 1990 |