Recall Z-1982-2019 ยท JQP Calculator/Data Processing Module, For Clinical Use

Beckman Coulter Inc.

Kaluza C Flow Cytometry Software, RX Only in the U.S.A., Beckman Coulter, Inc. for the following software devices: (A) Kaluza C Perpetual License (B) Kaluza C Single Use 1 Year License (C) Kaluza C 5 User Network (D) Kaluza C 10 User Network (E) Kaluza C Perpetual Educational License (F) Kaluza C Single User 1 Year educational License (G) Kaluza C 10 User Network Educational License (H) Kaluza C

DateSep 17, 2018
StatusTerminated
ReasonCompatibility issue between the software and Microsoft updates to Windows 7, 8, and 10. The compatibility issue causes the software to be unusable which could result in a delay of reporting results.
Root causeSoftware Manufacturing/Software Deployment
ActionAn "Urgent Medical Device Recall" letter was sent on September 18, 2018 via email to all affected customers. Beckman Coulter ask that if you are using a LIS, please review all previous LIS transmissions which included keywords for accuracy by comparing keywords displayed in Kaluza C Information Plots to those in the LIS system. Please report any erroneous transmissions to Beckman Coulter. The recall letter also ask their customers to visit the Beckman Coulter website to download version Kaluza C 1.1 and install on all affected systems. Installation of Kaluza C version 1.1 will prevent softwa
Quantity41 units
DistributionUS: CA and NY OUS: United Kingdom, Switzerland and Germany
FDA recordOpen on fda.gov