ClearanceBrief
Recall Z-1878-2019 ยท DSA Cable, Transducer And Electrode, Patient, (Including Connector)

GE Healthcare, LLC

Leadwire sets with snap ends: REF/Catalog Number Description 2106381-001 ECG Leadwire set, 5-lead, snap, AHA, 74 cm/ 29 in 2106381-002 ECG Leadwire set, 5-lead, snap, AHA, 130 cm/ 51 in 2106381-003 ECG Leadwire set, 5-lead, snap, IEC, 74 cm/ 29 in 2106381-004 ECG Leadwire set, 5-lead, snap, IEC, 130 cm/ 51 in 2106381-005 ECG Leadwire set, 5-lead, snap, AHA, mix 74 cm/ 29 in, 130 cm/ 51 in 2106383-

DateMay 10, 2019
StatusTerminated
ReasonECG trunk cables and leadwires may reduce the amount of energy reaching the patient during defibrillation, potentially limiting success of defibrillating the patients rhythm.
Root causeProcess design
ActionThe firm, GE Healthcare, initiated a "URGENT MEDICAL DEVICE CORRECTION"letter dated 05/10/2019. The letter explained the reason for recall and provided the following safety instructions: The affected ECG trunk cables and leadwires may continue to be used for monitoring only. Discontinue use of the affected ECG trunk cables and leadwires for patients where an arrhythmia that might require defibrillation is foreseeable, and in these patients use unaffected ECG trunk cables and leadwires. If defibrillation turns out to be unexpectedly needed when the affected ECG trunk cables and leadwires are
Quantity58,045 units (31,108 US; 26937 OUS) in total
DistributionWorldwide Distribution: US (nationwide) including Puerto Rico, and countries of: Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Bermuda, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvado
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.