Recall Z-1906-2019 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Bio-Rad Laboratories, Inc

EVOLIS Microplate System (Catalog no. 89700) with TimeLiner Data Management Software, (Catalog no. 90471A.

DateJan 10, 2019
StatusOpen, Classified
ReasonUser-induced circumstances can contribute to the EVOLIS Microplate Processor System not handling repeat testing requests/results properly and may contribute to transmitting erroneous results to the customer's Laboratory Information System (LIS).
Root causeSoftware design
ActionThe recalling firm issued letters dated 1/9/2019 via FedEx on 1/10/2019 to all EVOLIS customers.
Quantity213 devices
DistributionDistribution was nationwide to medical facilities, including to Puerto Rico. There was also government/military distribution. There was no foreign distribution.