Recall Z-1771-2019 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Philips North America, LLC
Philips Patient Information Center iX configured with a regional setting which uses a comma for a decimal separator: 866389 Patient Information Center iX B.01 (or later); 866390 Patient Information Center iX B.01 upgrade (or later) Product Usage: The intended use of the Information Center Software is to display physiologic waves, parameters, and trends, format data for strip chart recordings and
| Date | Apr 29, 2019 |
| Status | Terminated |
| Reason | Infusion Pump Data Storage Accuracy-Data from the BBraun Space LAN or Arcomed UniqueDoc infusion pumps transmitted via the HL7 output interface through the Patient Information Center iX using a LAN driver may be recorded in the patients chart or electronic medical record at exactly 100 times the actual bolus rate, infusion rate and total volume values |
| Root cause | Software design |
| Action | Philips initiated recall by letter on 4/29/19 identifies the reason for recall, helath risk and to follow the Action to be Taken by Customer/User section of the Customer Information Letter: Users are to be aware that the values in the patients chart or electronic medical record (EMR) may be in error and to confirm infusion pump parameters at the respective pump and only use these values to inform orders or to modify the infusion rate. Philips will provide a software update to customers using the BBraun Space LAN or Arcomed UniqueDoc infusion pumps, at no charge. For further information or su |
| Quantity | 12,391 |
| Distribution | Worldwide Distribution - US Nationwide Foreign: Andorra Aruba Australia Austria Azerbaijan Bahrain Bangladesh Belarus Belgium Bolivia Brazil Bulgaria Canada Chile China Colombia Costa Rica Croatia Czech Republic Denmark Dominican Rep Egypt Estonia Finland France French Guiana Germany Ghana Greece Gu |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K153702 | M3290B Philips IntelliVue Information Center iX Philips Medical Systems, Jun 13, 2016 |