ClearanceBrief
Recall Z-1940-2019 ยท DSP System, Balloon, Intra-Aortic And Control

Datascope Corp.

Cardiosave Rescue IABP, 0998-00-0800-83 Intra-Aortic Balloon Pump For cardiovascular use.

DateJul 16, 2019 (initiated Jun 17, 2019)
StatusOpen, Classified
ReasonIf battery maintenance is not performed appropriately, the battery may provide less than the minimum run time of operating power. Revised labeling and training are being provided.
Root causeUse error
ActionOn June 17, 2019, the firm notified its customers of the product issue via Urgent Medical Device Correction letters. Customers are to take the following immediate interim actions: " Ensure the IABP is plugged into an AC power outlet whenever possible during patient use to prevent the battery from depleting. " Ensure the IABP is plugged into an AC power outlet when the system is not in use. The batteries should be kept at a full charge even when the IABP is not in use. " When transporting patients within or between facilities, please refer to the IABP Operating Instructions Manual for recomme
Quantity6,197
DistributionWorldwide distribution. US Nationwide, AFGHANISTAN , HUNGARY , PANAMA , ALGERIA , ICELAND , PARAGUAY , ARGENTINA , INDIA , PERU , AUSTRALIA , INDONESIA , PHILIPPINES , AUSTRIA , IRAN , POLAND , AZERBAIJAN , IRAQ , PORTUGAL , BAHRAIN , IRELAND , QATAR , BANGLADESH , ISRAEL , ROMANIA , BELARUS , ITALY
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K112372CARDIOSAVE HYBRID INTRA-AORTIC BALLOON PUMP
Cardiac Assist, Maquet Cardiovascular, LLC, Sep 15, 2011
K151254CARDIOSAVE Hybrid Intra-Aortic Balloon Pump
Datascope Corp., Jul 2, 2015
K163542CARDIOSAVE Hybrid and Rescue Intra-Aortic Balloon Pump, CARDIOSAVE Li-Ion Battery Transport and Storage Case
Datascope Corp., Jan 31, 2017
K181122CARDIOSAVE Intra-Aortic Balloon Pump
Datascope Corp., May 31, 2018