ClearanceBrief
Recall Z-1715-2019 ยท LKK Pump, Infusion, Implanted, Programmable

Flowonix Medical Inc

Prometra II Programmable Pump, Cat. No. 13827 For intrathecal infusion.

DateMar 21, 2019
StatusTerminated
ReasonThe pump shuts down unexpectedly if Error 115, an alarm associated with a watchdog timer timeout, is triggered. There has been an increase of Error 115 occurrences due to a software defect caused by a Healthcare Provider using a Clinician Programmer, Software Version 2.00.29, to program a Bridge or Demand bolus while the pump is set to either a Periodic Flow or Multiple Rates flow mode. Error 115 can also occur if dendrites form or if the pump is dropped.
Root causeSoftware change control
ActionOn March 21, 2019, the firm issued Urgent Recall Notices to customers to inform them of the product issue. In May 2019, the firm sent updated communications to customers: Healthcare providers who have Clinician Programmers with SW 2.00.29 were sent a follow-up recall letter. This letter included the information that the pump stop following "Error Code 115" will only occur after the Bridge or Demand Bolus is complete, which could take up to seventy-two (72 hours). In addition, in the event of any cause of an "Error Code 115", the pump sounds a one-beep alarm and then shuts down. This behav
Quantity6445
DistributionUS nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.