Recall Z-1705-2019 ยท DXC Clamp, Vascular

Edwards Lifesciences, LLC

IntraClude Intra-Aortic Occlusion Device, REF: ICF100, 10.5 Fr (3.5 mm) X 39.4" (100 cm), Rx Only, Sterile EO The IntraClude intra-aortic occlusion device is indicated for use in patients undergoing cardiopulmonary bypass. The IntraClude intra-aortic occlusion device occludes and vents the ascending aorta when the balloon is inflated.

DateJun 11, 2019 (initiated May 14, 2019)
StatusTerminated
ReasonPossibility of balloon rupture in intra-aortic occlusion devices. If the balloon bursts during cardiopulmonary bypass, the heart can fill and warm, the operative site may be obscured, and the device will need to be exchanged or operative strategy would need to change, including placement of an external cross-clamp, conversion to an open procedure, or performing the procedure under fibrillation.
Root causeDevice Design
ActionOn 05/14/19, the firm, Edwards, mailed "Urgent Product Recall" notices, via overnight service to its customers. The letter described the product, problem and actions to be taken. The letter informed customers of the recall and instructed them to do the following: 1) Review this field safety notice to understand the potential hazard. 2) Complete and return the attached Acknowledgement Form within five (5) business days of receiving this notice to Customer Service. 3) Complete and return the Product Reconciliation Form to Customer Service; return affected devices with the Return Goods Authorizat
Quantity6552
DistributionWorldwide Distribution: US (nationwide) to states of: UT, GA, NJ, CA, OH, PA, AL, FL, TN, TX, ME, MA, IN, MS, NY, VA, MI, IL, CO, OR, WA, SC, AZ, SD, MO, CT, OK, NV, WV, WI, NC, MN, KY; and countries of: United Arab Emirates, Austria, Australia, Belgium, Canada, Switzerland, Colombia, Czech Republic
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K113182EDWARDS LIFESCIENCES INTRACLUDE INTRA-AORTIC CATHETER
Edwards Lifesciences, LLC, Dec 7, 2011
K132175EDWARDS LIFESCIENCES INTRACLUDE INTRA-AORTIC CATHETER
Edwards Lifesciences, LLC, Sep 13, 2013
K163693IntraClude Intra-Aortic Occlusion Device
Edwards Lifesciences, LLC, Jan 26, 2017