Recall Z-1893-2019 ยท MLZ Vitrectomy, Instrument Cutter

Johnson & Johnson Surgical Vision Inc

Disposable 25GA Vitrectomy Cutter, REF NGP0025, RX Only, Abbott Medical Optics Inc., The product provides the item necessary to perform one vitrectomy procedure. It is used to remove vitreous from the eye. It has been designed as a disposable (single use)

DateSep 24, 2018
StatusTerminated
ReasonPackaging error, 20GA Vitrectomy Cutter may be found in a 25GA package. Use of a Vitrectomy cutter that is a different size than expected could lead to the need to alter the surgical technique, including corneal incision enlargement.
Root causePackaging process control
ActionThe firm, Johnson & Johnson Vision, sent an"URGENT MEDICAL DEVICE RECALL" Letter dated September 20, 2018 overnight on September 20, 2018 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: If you have received impacted product, please immediately take the following actions: 1. Review your inventory to identify units of affected lot; 2. Stop using and remove from your inventory all affected lot; and 3. regardless if you do not have inventory, please complete the Customer Reply form and return product. If
Quantity392 units
DistributionUS Distribution to states of: NC, NV, TX, NY, UT, C, MN, NY, WA, NJ, OH, NV, CA and Internationally to: Australia and Japan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081681VITRECTOMY CUTTER AND ACCESSORIES
Advanced Medical Optics, Inc., Oct 8, 2008