Recall Z-2357-2019 ยท NBW System, Test, Blood Glucose, Over The Counter

Cambridge Sensors Limited

Microdot Xtra blood glucose test strips

DateJul 26, 2018
StatusOpen, Classified
ReasonThe firm reported that a faulty cutting/punching machine did not meet specifications after being fitted with new blades and did not correctly "notch" BG test strips intended for use in the US. The impacted strips did not fit into the inlet port of the BG meter, therefore resulting in a delay in BG testing and results.
Root causeNonconforming Material/Component
ActionOn July 26, 2018, the firm sent an email to it's US distributor that requested a sub-recall.
Quantity17,784 (USA)
DistributionIllinois
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K083273MICRODOT XTRA
Cambridge Sensors Limited, Nov 19, 2009