Recall Z-1496-2019 ยท IMD Pack, Hot Or Cold, Disposable

PF Consumer Healthcare 1 LLC

ThermaCare HEATWRAPS, ADVANCED BACK PAIN THERAPY, SKU: a) F00573301020A b) F00573301026X c) F00573301009B Product Usage: ThermaCare(R) HeatWraps are indicated in adults to provide heat therapy for the temporary relief of minor muscular and joint pains associated with overexertion, strains, sprains, and arthritis and menstrual pain. ThermaCare HeatWraps are made up of heat cells enclosed within wr

DateApr 26, 2019
StatusTerminated
ReasonThere is the potential that a HeatWrap could include cells that have a higher cell temperature than specified in lot S97473. The use of heat cell wrap with increased temperatures poses a potential risk of skin injuries such as burns/blisters and/or skin irritation on the wrap-applied area.
Root causeOther
ActionThe firm initiated the recall by press release on 04/26/2019. Notification packages to direct consignees were sent on 04/29/2019. The consignees were directed to cease distribution and return the product. Distributors were directed to sub-recall. Retailers were asked to post the press release either at the cash register or another prominent location.
Quantity225617 individual heat wraps
DistributionWorldwide distribution - US Nationwide and PR, and countries of Ireland, UK, Malta, Netherlands, Germany, and Switzerland.
FDA recordOpen on fda.gov