Recall Z-1672-2019 ยท DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass

Maquet Cardiovascular, LLC

Maquet Getinge-BEQ-TOP 35503 SGD Pack Material:701068797R01

DateMay 31, 2019 (initiated Apr 18, 2019)
StatusTerminated
ReasonCustom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured
Root causeDevice Design
ActionMaquet Getinge issued recall notification letter on April 18, 2019 via FedEx 2-day delivery identifies issue, health risk and action to take: quarantine and return to Maquet/Getinge recalled product. Complete response form and return to acknowledge this recall. Questions contact Getinge Customer Service at 1 888 627 8383 (press option 2, then option 2) between the hours of 8 AM and 6 PM Eastern Standard Time. On 3/26/20 firm expanded recall to include an additional lot for one consignee,product: "BEQ TOP 25003 3/8 3/8 up to 7 lpm", item number 701063255R02, lot 3000087051.
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K080592HLM TUBING SET WITH BIOLINE COATING
Maquet Cardiopulmonary, AG, Sep 11, 2008