Recall Z-1669-2019 ยท DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass
Maquet Cardiovascular, LLC
Maquet Getinge- (1) Material :701067370R01 BO-TOP 38803 OPEN HEART PACK W (2) Material: 701067370R02 BO-TOP 38803 OPEN HEART PACK W (3) Material: 701067370R03 BO-TOP 38803 OPEN HEART PACK W
| Date | May 31, 2019 (initiated Apr 18, 2019) |
| Status | Terminated |
| Reason | Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured |
| Root cause | Device Design |
| Action | Maquet Getinge issued recall notification letter on April 18, 2019 via FedEx 2-day delivery identifies issue, health risk and action to take: quarantine and return to Maquet/Getinge recalled product. Complete response form and return to acknowledge this recall. Questions contact Getinge Customer Service at 1 888 627 8383 (press option 2, then option 2) between the hours of 8 AM and 6 PM Eastern Standard Time. On 3/26/20 firm expanded recall to include an additional lot for one consignee,product: "BEQ TOP 25003 3/8 3/8 up to 7 lpm", item number 701063255R02, lot 3000087051. |
| Distribution | Nationwide |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K080592 | HLM TUBING SET WITH BIOLINE COATING Maquet Cardiopulmonary, AG, Sep 11, 2008 |