Recall Z-1786-2019 ยท NDC Calculator, Drug Dose

Monarch Medical Technologies

Monarch Medical Technologies EndoTool IV, Versions 1.8, Version 1.8.5 and higher, and Version 1.9 and Version 1.10.

DateMay 3, 2019
StatusOpen, Classified
ReasonInsulin dosing calculations were erroneously high.
Root causeSoftware change control
ActionThe firm notified the consignees by email beginning on 05/03/2019. The notice alerted the consignee of the problem and provided an educational brief and a tip card to provide to the staff to remind them to always use clinical judgement in following EndoTool IV dose recommendations.
Quantity83 downloads
DistributionNationwide and Dubai
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K132547ENDO TOOL GLUCOSE MANAGEMENT SYSTEM (EGMS)
Monarch Medical Technologies, Apr 17, 2014