Recall Z-1955-2019 ยท NIE Catheter, Hemodialysis, Triple Lumen, Non-Implanted

COVIDIEN LLC

MAHURKAR and Argyle acute hemodialysis catheters, Material numbers 8813793009 8813793013 8813794005 8813794009 8813816005 8813816009 8813817005 8813817009 8815668007 8815668011 8815668021 8815668027 8817142005 8817143005 8817145005 8817146001 8817146007 8817149007 8817232018 8817232019 8817277007 8817277011 8817277021 8817277027 8830414001 8830414002 8830415001 8830415003 8830416001 8830416003 883

DateApr 29, 2019
StatusTerminated
ReasonThe priming volume values printed on the MAHURKA and Argyle acute catheters and Instructions for Use (IFU) are higher than the volumes required to fill each lumen.
Root causeError in labeling
ActionOn April 29, 2019, consignees were notified by letter via Federal Express and/or federal mail. The Field Safety Alert informed customers that the priming volume values printed on the MAHURKAR and Argyle acute catheters and Instructions for Use are higher than the volumes required to fill each lumen. The firm will update the IFU for the affected products to provide clarity on the printed priming volumes. In the meantime, the firm recommends the use of a non-heparinized lock solution, such as 4% sodium citrate or 1000 U/mL heparin to mitigate potential risk associated with unintended administra
Quantity2,223,067
DistributionNationwide domestic distribution. Foreign distribution to Canada, Australia, New Zealand, Latin America, and Hong Kong.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K020089MAHURKAR TRIPLE LUMEN CATHETER, 12 FR
The Kendall Company, Apr 10, 2002
K030209MAHURKAR OPLUS CATHETER, MODEL 13.5 FR
The Kendall Company, May 13, 2003
K092797ACUTE AND CHRONIC CATHETER REPAIR KIT
Kendall, Nov 19, 2009
K102605MAHURKAR TRIPLE LUMEN DIALYSIS CATHETER
Covidien, LLC, Formerly Registered AS Tyco Healthca, Dec 22, 2010
K893188QUINTON SINGLE LUMEN CATHETER
Quinton, Inc., Jun 8, 1989