Recall Z-1369-2019 ยท KOB Catheter, Suprapubic (And Accessories)

Cook Inc.

Cook Kwart Retro-Inject Ureteral Stent Set: PRODUCT/CATALOG NUMBER/GPN: 4.7 Fr 003500 G14916; 6.0 Fr 003600 G14836; 7.0 Fr 003700 G14837; 8.0 Fr 003800 G14844; Set with AQ Hydrophilic Coating PRODUCT/CATALOG NUMBER/GPN: 4.7 Fr AQ-003500 G17150; 6.0 Fr AQ-003600 G17151; 7.0 Fr AQ-003700 G17152

DateApr 22, 2019
StatusTerminated
ReasonKwart RetroInject Stent Set (Multi-Length) and Salle Intraoperative Pyeloplasty Stent Set Instructions for Use (IFU) did not include a warning for stent knotting.
Root causeError in labeling
ActionCook issued MEDICAL DEVICE CORRECTION letters via courier to US Consignees on April 22, 2019. The letter provides reason for recall, products affected, health risk, action to take: Understand that stent knotting is a potential complication associated with use of the Kwart Retro-Inject" Stent Set and Salle Intraoperative Pyeloplasty Stent Set and should be considered if significant resistance is encountered during attempts at removal. 2. Please maintain a copy of this notice with the current IFU or product(s) in your inventory. 3. If you have affected product in inventory, you may continue to u
Quantity12968
DistributionNationwide Foreign: Australia Canada Austria Belgium Denmark Germany Spain Finland France Martinique French Polynesia Switzerland Ireland Italy Netherlands Poland Sweden United Kingdom Hong Kong New Zealand Chile Colombia Israel Russian Federation Czech Republic Greece Jordan Slovakia South Africa U