Recall Z-0025-2020 ยท EOQ Bronchoscope (Flexible Or Rigid)

AURIS HEALTH INC

Monarch, Aspirating Biopsy Needle, REF: MBR-000012, Rx Only, UDI: B634MBR0000120

DateApr 19, 2019
StatusTerminated
ReasonReports of aspirating biopsy needle breakage
Root causeLabeling design
ActionOn 04/19/19, Urgent Medical Device Notifications were mailed to customers. Customers were advised the following: Please review the updated instructions for use for proper insertion, along with updated and warnings/cautions related to the use of the device. Specifically, the following instructions were included: " Do NOT use excessive insertion force if resistance is encountered during insertion of the needle through the bronchoscope working channel. Excessive force may cause damage to either the device or the bronchoscope working channel. " DO inspect the needle sharp for damage prior to use.
Quantity158
DistributionUS nationwide distribution to the states of: CT, FL, MI, OK, TN, MS, CA, MN, OH, PA, and IL.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K173760Monarch Endoscopy Platform (Monarch Platform)
Auris Surgical Robotics, Inc., Mar 22, 2018