Recall Z-1742-2019 ยท JSO Culture Media, Antimicrobial Susceptibility Test, Excluding Mueller Hinton Agar

Bio-Rad Laboratories, Inc

MRSASelect, IVD, REF 63747, containing 20 plates per package. The MRSASelect is indicated for use in conjunction with other laboratory tests and clinical data available to aid in the identification and diagnosis of MRSA from patients with skin and soft-tissue infections.

DateJun 8, 2018
StatusTerminated
ReasonAbnormal coloration of agar plates and the growth of non-characteristic MRSA colonies.
Root causeNonconforming Material/Component
ActionThe firm. Bio-Rad laboratories, Inc., issued an "URGENT PRODUCT CORRECTION" letter dated 6/7/2018 to all customers via FedEx on 6/7-8/2019. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: " As already recommended in the package insert, please ensure that exposure of agar plates to light is minimized, both before and during incubation, with the added time restriction of no more than 2 hours during storage, warming, incubation, and reading. " Do not use agar with a change in color from white to an orange-pinkish shade. " As re
Quantity5,212 units
DistributionUS Nationwide Distribution to: AK, AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, and Puerto Rico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081212MRSASELECT - EXTENDED INCUBATION