Recall Z-1690-2019 ยท MQV Filler, Bone Void, Calcium Compound

Exactech, Inc.

Opteform Allograft Disc, 90mm x 5mm, 22cc, Catalog Number 600-03-90

DateApr 26, 2019
StatusTerminated
ReasonThe Opteform Disc-30mm, Opteform Disc-45mm, and Opteform Disc-90mm were potentially exposed to higher temperatures than those documented in the IFU due to a failure of the cold storage equipment.
Root causeEnvironmental control
ActionThe firm initiated the recall on 04/26/2019 by electronic mail. The notice requested the consignee cease distribution of the product, notify their customers (the user), quarantine all units subject to recall in inventory, and if any units were implanted to follow-up on the patient's condition 30 days after surgery.
Quantity3 units
DistributionOK, VA, FL
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K043421OPTEFORM, OSTEOFIL, RTI ALLOGRAFT PASTE IC, RTI ALLOGRAFT STRIP IC
Regeneration Technologies, Inc., Feb 16, 2005